Measured per NU-1, 1994 using Tc-99m where applicable.
| Performance characteristic | Nucline™ X-Ring | Units | |||||
|---|---|---|---|---|---|---|---|
SPECT UFOV (Useful Field of View) | 215 x 230 | mm | |||||
Number of PMTs | 4 x 60 | 4 x 33 |
| ||||
Crystal thickness | 6.4 | 9.5 | 12.5 | mm | |||
Intrinsic spatial resolution | CFOV | FWHM | 2.3 | 2.9 | 3.4 | 3.9 | mm |
|
| FWTM | 4.6 | 5.8 | 6.8 | 7.8 | mm |
| UFOV | FWHM | 2.4 | 3.0 | 3.5 | 4.0 | mm |
|
| FWTM | 4.8 | 6.0 | 7.0 | 8.0 | mm |
Intrinsic energy resolution | UFOV | FWHM | 9.7 |
| |||
Intrinsic flood field uniformity | CFOV | Differential | 1.9 | % | |||
|
| integral | 2.4 | % | |||
| UFOV | differential | 2.4 | % | |||
|
| integral | 2.9 | % | |||
Intrinsic spatial linearity | CFOV | differential | 0.18 | mm | |||
|
| absolute | 0.38 | mm | |||
| UFOV | differential | 0.20 | mm | |||
|
| absolute | 0.40 | mm | |||
Intrinsic count rate performance without scatter | maximum | 200 | kcps | ||||
at 20% loss | 160 | kcps | |||||
System spatial resolution with LEGP / LEHR collimator | FWHM | 8.3 / 6.6 | 8.5 / 6.9 | 8.7 / 7.1 | 8.9 / 7.2 | mm | |
FWTM | 16.1/13.1 | 16.5/13.5 | 16.7/13.7 | 16.9/13.9 | mm | ||
Maximal matrix size | 1024 x 1024 | pixel | |||||
Camera weight | 800 | kg | |||||
Minimal room size | 12 | m2 | |||||
Uptime warranty | 98 | % | |||||
Movements directions | 2 |
| |||||
Collimator mounting | slide-in, bolt fixing |
| |||||
Collimator changing time | < 1.5 | min | |||||