Measured per NU-1, 1994 using Tc-99m where applicable.
Performance characteristic * | Nucline ™ | |||||
|---|---|---|---|---|---|---|
Type ( Number of PMTs / Useful Field of View ) | TH 22 ( 22 / 180 x 180 )TH 33 ( 33 / 230 x 215 )TH 45 ( 45 / 270 x 270 ) |
| ||||
Crystal thickness | 6.4 | 9.5 | 12.5 | mm | ||
Intrinsic spatial resolution | CFOV | FWHM | 2.9 | 3.6 | 4.1 | mm |
|
| FWTM | 5.8 | 7.2 | 8.2 | mm |
| UFOV | FWHM | 3.0 | 3.7 | 4.2 | mm |
|
| FWTM | 6.0 | 7.4 | 8.4 | mm |
System spatial resolution with LEAP / LEHR collimator | FWHM | 8.5 / 7.0 | 8.7 / 7.2 | 8.9 / 7.4 | mm | |
FWTM | 16.5 / 13.5 | 16.7 / 13.7 | 16.9 / 13.9 | mm | ||
System sensitivity | 240 / 150 | 260 / 160 | 280 / 180 | cpm/µCi | ||
Intrinsic energy resolution | UFOV | FWHM | 9.9 | % | ||
Intrinsic flood field uniformity | CFOV | differential | 1.9 | % | ||
|
| integral | 2.4 | % | ||
| UFOV | differential | 2.4 | % | ||
|
| integral | 2.9 | % | ||
Intrinsic spatial linearity | CFOV | differential | 0.18 | mm | ||
|
| absolute | 0.38 | mm | ||
| UFOV | differential | 0.20 | mm | ||
|
| absolute | 0.40 | mm | ||
Intrinsic count rate performance without scatter | maximum | 200 | kcps | |||
at 20% loss | 160 | kcps | ||||
Maximal matrix size | 1024 x 1024 | pixel | ||||
Camera weight | 180 | kg | ||||
Minimal room size | 6 ( 2 x 3 ) | m2 | ||||
Uptime warranty | 98 | % | ||||
On wheel movement | Yes |
| ||||
Movements directions | 3 |
| ||||
Collimator fixing | 2 directional auto-fixing |
| ||||
Collimator changing time | < 0.2 | min | ||||