:: PRODUCTS ::
Measured per NU-1, 1994 using Tc-99m where applicable.
Performance characteristic |
| ||||
|---|---|---|---|---|---|
Type | R | R-HR | Units | ||
UFOV (Useful Field of View) | 530 x 390 | mm | |||
Intrinsic energy resolution | UFOV | FWHM | 10.2 | 9.7 | % |
Intrinsic spatial resolution | CFOV | FWHM | 4.0 | 3.4 | mm |
|
| FWTM | 8.0 | 6.8 | mm |
| UFOV | FWHM | 4.1 | 3.5 | mm |
|
| FWTM | 8.2 | 7.0 | mm |
Intrinsic flood field uniformity | CFOV | differential | 1.9 | 1.9 | % |
|
| Integral | 2.4 | 2.4 | % |
| UFOV | differential | 2.4 | 2.4 | % |
|
| Integral | 2.9 | 2.9 | % |
Intrinsic spatial linearity | CFOV | differential | 0.18 | 0.18 | mm |
|
| absolute | 0.38 | 0.38 | mm |
| UFOV | differential | 0.20 | 0.20 | mm |
|
| absolute | 0.40 | 0.40 | mm |
Intrinsic count rate performance without scatter | maximum |
| 240 | 240 | kcps |
at 20% loss |
| 160 | 160 | kcps | |
System spatial resolution with LEGP/LEHR collimator | FWHM | 9.0 / 7.5 | 8.8 / 7.3 | mm | |
FWTM | 17.2 / 14.4 | 16.8 / 13.9 | mm | ||
Center of rotation | 0.2 ( 64 x 64 matrix ) | pixel | |||
Maximal matrix size | 1024 x 1024 |
| |||
Collimator changing time | < 0.5 | min | |||
Camera weight with table | 1350 | kg | |||
Minimal room size | < 18 ( 3.8 x 4.7 ) | m2 | |||
Uptime warranty | 98 | % | |||
© 2004-2005 Mediso Ltd. All rights reserved.